| AMOUNT OF SUBSTANCE (Inorganic analysis) | ||||||||||||
|
Quantity - Amount of substance |
||||||||||||
|
||||||||||||
| Tasks | ||||||||||||
|
The main task is the realisation of base unit mole and its transfer into reference materials. |
||||||||||||
| International comparisons | ||||||||||||
|
The accuracy of measurement is substantiated by comparisons organised by BIPM (CCQM-P7, CCQM-P19.1, CCQM-K8, CCQM-P36, CCQM-K42, CCQM-K29, CCQM-K59). |
||||||||||||
| Available metrological services | ||||||||||||
| Certified reference materials (CMRs) of chemical composition - pure substances with expanded uncertainty 0.004 % to 0.012 % and calibration solutions with expanded uncertainty 0.2 % to 0.3 %. | ||||||||||||
Methods of measurement: |
||||||||||||
|
- the amount content
in primary CRMs is certified by high-accuracy coulometric titration. - the mass concentration of monoelemental or anion calibration solutions is certified by refined methods of classical analysis traceable to primary CRMs or directly on the National Standard. |
||||||||||||
| Instruments in use | ||||||||
![]() |
High accuracy coulometric equipment, declared as National Standard for Amount of Substance. Measuring range is 0.0001 mol to 0.01 mol. The measurements are traceable to the value of the Faraday constant and Slovak standards of electric resistance, voltage, time and mass. | |||||||
|
||||||||
| |
||||||||||
|
||||||||||
| Quantity - pH. |
||||||||||
| Unit
- dimensionless. |
||||||||||
| Tasks | ||||||||||
|
The main task is the realisation of the unit of pH. The unit is further transfered by means of Certified Reference Materials. |
||||||||||
| International comparisons | ||||||||||
| The
accuracy of measurement substiated organised by BIPM (CCQM-K9, CCQM-K17,
CCQM-P37, CCQM-P52, CCQM-K18) and organised by COOMET 322/Ru/04.
|
||||||||||
| Available metrological services | ||||||||||
|
Establishing of traceability of pH by the means of CRMs. Production and distribution of primary cerified reference materials - buffers pH range from 1 to 10 with expanded uncertainty 0.005 - 0.02. |
||||||||||
Methods of measurement: |
||||||||||
| -
primary buffers are
certified by definition method in Harned's cell, - secondary CRMs of I. order are traceabled to primary CRMs by measurement in the definition cell with two hydrogen electrodes, - secondary CRMs of II. order are traceabled to primary CRMs by measurement in the cell with transfer through a glass electrode. |
||||||||||
| Traceability is disseminated by means of primary CRMs in the following nominal values: 1.679; 4.000; 7.000; 9.180; 10.000. | |||||||||||||
| Instruments in use | |||||||||||||
![]() |
|
||||||||||||
|
|||||||||||||
| |
||||||||||
|
||||||||||
| Quantity
- electrolytic conductivity. |
||||||||||
|
Unit - the unit of electrolytic conductivity is S/m (Siemens per meter). It is defined as the conductivity between two oposite faces of the cube with 1 m edge, filled with the measured solution. |
||||||||||
| Tasks | ||||||||||
|
The
main tasks are maintenance, improvement of the realisation and
dissemination of the unit of electrolytic conductivity, realisation of
the primary standard and certification of the primary Certifed Reference
Materials (CRMs) as a means for certification of secondary CRMs.
Traceability chain is realised by means of CRMs |
||||||||||
| International comparisons | ||||||||||
|
2001 -
Trilateral measurement comparison - PTB (DE), SMU (SK), UkrCSM (UA);
pilot SMU. 2003 - CCQM-P47 (pilot NMi) 2005 - CCQM-K36 (pilot DFM) |
||||||||||
| Available metrological services | ||||||||||
| Production of electrolytic conductivity CRMs with following nominal values: 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1 and 2 S/m. Calibration of conductivity cells. Calibration of solutions supplied by the customer | ||||||||||
Methods of measurement: |
||||||||||
| The measurement system for conductivity is based on the comparison bridge method in a two-electrode Jones cell. The primary RMs (water solutions of KCI) are measured by means of primary standard od electrolytic conductivity in the measuring range of 0.01 - 2 S/m at the temperature of 25.00°C. | |||||||||
| Instruments in use | |||||
![]() Conductivity cell with exchangeable central part |
Measurement of electrolytic conductivity is performed in a cell with exchangeable central tube whose length and cross-section are known. Conductivity is measured by means of an alternating current at a 1000 Hz frequency | |||||||||||||||
| Metrological parameters: conductivity range is 0.01 S/m - 2 S/m; relative standard uncertainty, type A is EA = 0.00006; relative standard uncertainty, type B is EB = 0.00023; relative combined standard uncertainty is EC = 0.00024. | ||||||||||||||||
|
||||||||||||||||