ACTIVITY OF RADIONUCLIDES LABORATORY
       
      Quantity - activity.
      Unit - Bq (becquerel).
      Tasks of the laboratory
      The main task of the Laboratory is the realisation of the unit of radionuclide activities and its transfer to user's instruments. Derived quantities used are mass activity, volume activity and surface activity in corresponding units.
      International comparisons
      CCRI (II) comparison of 152Eu activity measurement (BIPM 1999-2000). DUNAMET D29 common calibration of ionizing chambers (CMI 2001).
    Available metrological services
      Radionuclides calibrators verification (mainly to nuclear medicine laboratories). Instruments are tested to fulfill STN 40 4014 standard requirements of 10 % expanded uncertainty in the range from 1 MBq up to 100 GBq.
      Gammaspectrometers verification. Semiconductor detectors used in gammaspectrometry are verified in according with IEC 973 standard.
      Verification of monitoring equipment in liquid effluents. tests of calibration and linearity are made in according with IEC 861 standard (15 % expanded uncertainty).
      Verification of surface contamination monitors. Alpha and/or beta surface contamination monitors are tested for 14C, 60Co, 90Sr, 137Cs, 204Tl and 241Am calibration in according with STN 35 6566 standard (25 % expanded uncertainty).
      Sealed sources standardization and certification. Sealed sources activity can be by 4 ionizing chamber measurement. The chamber has bee traced directly to some PTB standards.
      Methods of measurement:
      - radionuclide calibrators are verified in situ with a travelling standard ionization chamber using radionuclides provided by users. Standard measurement conditions are given using standard containers (injection ampoules or peniciline vials) filled by 2 - 5 ml of radionuclide solution;
- gammaspectrometers are verified by calibrated water solutions of several radionuclides sent to users. Liquid effluent monitors are verified by calibrated 137Cs low-activity water solutions and a set of 137Cs solid standards. Contamination monitors are tested by certified large area sources checked by a standard instrument.
       
    Instruments in use
      Basic instrument for activity measurement is calibrated 4 (re-entrant) pressurized ionizing chamber. Low gama activity measurements and radionuclide purity testing are made by calibrated gammaspectrometric system with HPGe detector. set of large area (110 x 150) sources is available together with a verification instrument with two thin-window proportional counters.
       
     
Contact