DOSIMETRY LABORATORY
     
Dosimetry of gamma radiation Dosimetry of neutron radiation X-ray kerma
       
     
DOSIMETRY OF GAMMA RADIATION
       
      Quantity - air kerma, dose equivalent and their rates.
      Unit - Gy (gray), Sv (sievert), Gy·h-1, Sv·h-1.
      Tasks
      The main tasks are realisation and transfer to the user's instruments of the scale of dosimetry quantities from low dose rates (background level) througth the medium dose rates (radiation protection leveles) to the high dose rate (therapy level). The instrument type tests and personal certification in dosimetry field is a further activity of the Laboratory.
      International comparisons
     

2000 – 2002             : BIPM.RI(I)-K1 key comparison o fair kerma for 60Co

2005-2008                : No. 813 (EUROMET.RI(I)-K1 and EUROMET.RI(I)-K4) key comparison of air kerma and absorbed dose to water

2005-2006                : COOMET 318 regional key comparison of air kerma for 60Co

2000 – 2003             : EA IR3 comparison – radiation protection dosemeters

2000 – 2003             : EA IR4 comparison – personal dose equivalent

2007                          : multilateral intercomparison in radiation protection

 (H*(10) a Hp(10))


    Available metrological services
      Radiation protection - photon rays: Therapy level - photon rays:
     
- calibration of dosimeters and reference standards,
- calibration of personal dosimeters,
- calibration of users reference fields,
- calibration of dosimeters and reference standards.
     
- tests and verification of dosimeters, personal dosimetry services.
      Click for details!
      Methods of measurement:
      - calibration against the calibrated reference photon field is the main method used. Also the substitution method is available. The reference beams are produced in accordance with ISO 4037-1 and ISO 4037-2;
- calibration and tests are performed in accordance with the working procedures of SMU based on ISO and IEC standards (ISO 4037-3, STN IEC 61017-1, STN IEC 61017-2, IEC 61526, ISO 14146);
- for determination of absorbed dose in water the conversion procedure from air kerma based on IAEA TRS 277 code of practice is used. The calibration itself is done in water phantom.
       
    Instruments in use
      Radiation protection range is covered by the set of secondary standard ionising chambers Two TEMA irradiators with 137Cs and 60Co radiation sources are used for generating of calibrated photon fields. 3-D positionings variation of source to detector distance in wide range from 0.5 m to 7 m. The PMMA slab phantom is used for personal dose equivalent Hp(10) measurement. In the gamma therapy range, the calibrations are performed in collimated 60Co reference photon field. The photon field is calibrated against the primary standard graphite cavity ionising chamber and Keithley 6517A electrometer.
     
Set of standard ionising chambers

T
he irradiation facility for the radiation protection level for 137Cs and 60Co photon reference beams
     
Contact
Dr. Jozef Dobrovodský Mr. Norman Durný
phone: + 421 2 60294 671 phone: + 421 2 60294 668
e-mail: dobrovodsky@smu.gov.sk e-mail: durny@smu.gov.sk

                 
     
DOSIMETRY OF NEUTRON RADIATION
                   
      Quantity - absorbed dose, absorb dose rate, dose equivalent and their rates.
      Unit - Gy (gray), Sv (sievert), Gy·h-1, Sv·h-1.
      Tasks
      Main tasks are realisation and transfer to the user's instruments of dosimetry quantities of neutron.
      International comparisons
      2001: EUROMET/DUNAMET D28 - Comparison of the National Standard of the Neutron Dose Equivalent.
    Available metrological services
      Radiation protection - neutron:
      - calibration of dosimeters and reference standards;
     
Parameter Range Expanded uncertainty (k=2)
ambient dose equivalent (1·10-6 ÷ 1·10-3) Sv 14 %
ambient dose equivalent rate (1·10-6 ÷ 4·10-4) Sv·h-1 14 %
      - calibration of personal dosimeters;
     
personal dose equivalent (1·10-6 ÷ 1·10-3) Sv 14 %
personal dose equivalent rate (1·10-6 ÷ 4·10-4) Sv·h-1 14 %
      - tests and verification of dosimeters, personal dosimetry services.
      Methods of measurement:
      - neutron reference fields in accordance with the ISO 8529 standard are generated in a centre of 10 x 10 x 8 m large room by the underground installed set of 241AmBe, 239PuBe and 252Cf radionuclide neutron sources and the neutron generator by 1011 s-1 output is under the construction.
    Instruments in use
      Neutron dose equivalent rate standard - consists of tree substandards:
      Bonner Spherical Spectrometer
- the set of different volume tissue eqivalent proportional counters with related electronics;

- the set of paired ionising chambers with electrometer;

- the Bonner sphere spectrometer with the spherical 3He counter and electronics.
         
     
Contact
Mr. Valent Jenis
phone: + 421 2 60294 664
e-mail: jenis@smu.gov.sk
     
     
X-RAY KERMA
       
      Quantity - air kerma, dose equivalent and their rates.
      Unit - Gy (gray), SV (sievert), Gy·h-1, Sv·h-1.
      Tasks
      The main tasks are realisation and transfer to the user's instruments of dosimetry quantities of X-rays of the following quantities: ISO 4037-1 Narrow spectra, ISO 4037-1 Wide spectra and IEC 61267 RQA. The quality of the X-ray beams in Mammography diagnostic level is a further activity of the Laboratory.
      International comparisons
      2001: EUROMET 526 - Calibration of dosimeters used in mammography with different X-ray qualities.
    Available metrological services Click for details!  
     
Air kerma and air kerma rate, ambient dose equivalent and ambient dose equivalent rate: Personal dose equivalent and personal dose equivalent rate: Quality beams measurement: air kerma and air kerma rate:
- calibration of dosimeters and reference standards. - calibration of personal dosimeters and personal dosimemeter systems. - calibration of dosimeters and reference beams.
                 
    Instruments in use
     

Nomex, PTW
Instrument for quality beams measurement Norimex, PTW: absorbed dose 300 nGy ÷ 500 mGy; dose rate is 50 nGy·s-1 ÷ 5 mGy·s-1; kV range is 50 kV ÷ 150 kV; mAs range is 0.5 mAs ÷ 999.9 mAs.
Set of IC: OMH ND 1000; ND 1001; Xradin A4.
Electrometers: DPI2; Keithley 6517A.
X-ray generator: MG324, Philips: continuous, kV range is 14 kV ÷ 320 kV, current range is 0.05 mA ÷ 22.5 mA; XHF 30R, Creos: pilsed 100 kHz, kV range is 50 kV ÷ 120 kV, mAs range is max 640 mAs.
         
                 
     