LOW ABSOLUTE PRESSURE LABORATORY
       
      Quantity - pressure and flow under vacuum and molecular flow conditions.
      Unit - Pa, Pa·m3·s-1.
      Tasks of the laboratory
      The main task of Laboratory is the realisation of a low absolute pressure, parcial pressures and small flows.
  Available metrological services
     
Calibration and Measurement Services Range Combined Standard Uncertainty
Calibration of vacuum gauges. 1 kPa - 10 mPa 0.1 % - 1.2 %
1 mPa - 10-7 Pa (2 - 8) %
Calibration of helium leaks. 1·10-2 - 1·10-10 Pa·m3·s-1 (4 - 8) %
Calibration of oriffices conductivity, artificial conductivity for molecular flow. 10 - 1·10-8 Pa·m3·s-1 (2 - 7) %
Measurement of small flow in condition of molecular flow. 10 - 10-5 Pa·m3·s-1 (2 - 7) %
    Methods of measurement:
      - method of dynamic expansion is used to increase the National Standard range to 10-6 to 10-7 Pa. The dynamic expansion is realized by flow through members with known conductivity.
       
  Instruments in use
      National Standard - Compress mercury vacuum gauge of McLeon type (range 1 kPa - 10 mPa, combined standard uncertainty from 0.1 % to 1.2 %), stainless steel construction (decreases error caused by electrostatic charge arising from mercury motion in a capillary), glass capillary d = 8 mm, 80 kg mercury, construction eliminates the Ishii effect. Mercury column height is red by telescope with readability 0.01 mm.
      Capacitive vacuum gauge with three sensors of 100 kPa, 10 kPa, 1 Pa.
      Apparatus for dynamic expansion in the range 10-2 to 10-6 Pa.
      Quadrupole mass spectrometer.
      Apparatus for glass pressure increase depends on time in known volume.
      Set of liquid flowmeter.
       
     