TIME AND FREQUENCY LABORATORY
       
      Quantity
      Unit
      Tasks of the laboratory
      International comparisons
    Available metrological services
      Calibration of time and frequency quantum standards: cesium, rubudium (R and S, C1-50, SVC-74). Calibration of oscillators with high-stable crystal units (1 MHz, 5 MHz, 10 MHz). Calibration of oscillator unit for time measuring systems. Calibration of generators, frequency synthesizers. Calibration of electronic counters up to fmax = 48 GHz. Calibration of receivers of standard signals for station DCF-77 and Loran-C, navigation satellite system GPS. Calibration of comparators and frequency multipliers 1 Mhz, 5 MHz, 10 MHz for time interval higher than 0.1 s. Calibration of electronic stop watches and devices for calibration of mechanical stopwatches. Calibration of electronic stop watches by ESW 2000.
      Above mentioned devices are calibrated with the best standard combined uncertainty of 3x10-14.
      Methods of measurement:
      - the method used for determination of a relative frequency error at nominal values 1 MHz, 5 MHz, 10 MHz is the frequency difference measurement by a frequency comparator in connection with a frequency deviation multiplier;
- t
he method of standard signal comparison at the time scale followed by evaluation of the time scale differencies is used for long-term standard tests;
- a
short term frequency instability - measurement of Allan variance for oscillators with nominal values 1 MHz, 5 MHz, 10 MHz is measured by using the frequency comparator in connection with the frequency deviation multiplier in the range of time interval from 0.1 s to 1000 s with uncertainty of measurement 2x10-13.
       
    Instuments in use
      The SMU time and frequency primary standards are based on the cesium beam frequency standard in accordance with definition of the International system of units SI. The SMU Time and Frequency Standard is regulary compared direct to the value of UTC in the BIPM Paris by means of satellite system GPS. The relative error of standard is continuosly kept at < 4x10-14.The measurement of time and frequency generator, time meters, meters of time interval and frequency is provided in a range from 1 ns and at the frequency range up to 48 GHz.
       
     